Catalog Number 1012463-23
Device Problem Physical Resistance (2578)
Patient Problem Prolapse (2475)
Event Date 02/03/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported resistance during advancement and patient effect appears to be related to circumstances of the procedure.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.The absorb device is currently not commercially available in the u.S; however, it is similar to a device sold in the us.
 
Event Description
It was reported that the patient presented with unstable angina.The procedure on (b)(6) 2017 was to treat a lesion located in the distal left anterior descending (lad) artery with soft calcification, long stenosis, ectatic vessel and 85% stenosis.Vessel sizing was performed with angiography and determined to be 3 mm in diameter.Predilatation was performed with two semi-compliant balloons (2.0x30 and 2.5x30) with residual stenosis reduced to less than 40%.After vessel preparation, a 3.0x23mm absorb was advanced to the distal part of the lesion with some difficulty in deliverability causing a plaque shift; however, there were no adverse patient effects.The procedure was successfully completed with 2 absorb overlapped to treat the plaque shift and more proximal area of the lesion.There was a good final patient outcome.There were no reported adverse patient effects or clinically significant delay in procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM
Type of Device
BIORESORBABLE DRUG ELUTING SCAFFOLD
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6418723
MDR Text Key70377745
Report Number2024168-2017-02323
Device Sequence Number1
Product Code PNY
UDI-Device Identifier08717648157035
UDI-Public(01)08717648157035(17)170710(10)6011361
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P150023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 03/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/10/2017
Device Catalogue Number1012463-23
Device Lot Number6011361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-