Engineering investigation: the device in question was not returned therefore a thorough investigation cannot be conducted.If the device had been returned the surface would have been evaluated to determine if the stent was crimped properly during manufacturing.When the stent is crimped on to the folded balloon the stent frame leaves impressions on the surface of the balloon.The impressions indicate if the stent was properly crimped on to the balloon.A full review of the catheter lot history records for the devices in question was performed.The records indicate that this lot of catheters passed atriums final lot qualification testing.This inspection requires that the catheter lot must pass the following: · ability of the stent and delivery system to be passed through the labeled introducer sheath.· ability to deploy the stent at nominal pressure (8atm).· ability to withdraw the deflated balloon catheter back through the labeled introducer sheath · ability of the delivery system to withstand 5 inflate/deflate cycles at the rated burst pressure (12atm) without leaks or failures.· balloon burst testing.The balloon must burst over the rated burst pressure specified on the label (12atm).Result: all (b)(4) quality inspection samples passed this final inspection without any non-conformances noted during the final lot qualification testing.This includes passing the device through the labeled introducer sheath without stent movement.Conclusion: based on the details of the event and the successful lot qualification test data, atrium can find no fault with the lot of stent delivery systems in question.The case details indicate that this complaint was possibly due to the improper prepping of the device prior to use.Clinical evaluation: there are several possibilities that can cause stent dislodgement including but not limited to manipulation of the stent prior to use.If personnel at the table were unfamiliar with the device and attempted to hand crimp or manipulate the product in any way during preparation, it could interrupt the integrity of the stent.The instructions for use (ifu) warns that special care must be taken not to handle or in any way disrupt the placement of the icast covered stent on the balloon.This is most important during catheter removal from packaging, placement over the wire guide and advancement through the introducer.
|