The customer contacted a (b)(4) customer care center (ccc) specialist.The ccc reviewed the data provided.The customer's quality control (qc) was in range at the time of the event.A (b)(4) customer service engineer (cse) was dispatched to the customer's site.The cse checked all ultrasonic voltages, resulting within specification.The cse verified the wash probe, sample probe and reagent 2 probe alignments and adjusted probes.The cse calibrated the heterogeneous module mixers.The cse found material in the wash station tubing 2 and 4.The cse replaced the wash stating tubing, primed and ran 5 repetitions precision for qc levels 1 and 2, resulting within range.The cause of the discordant, discordant, falsely elevated free thyroxine result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
A discordant, falsely elevated free thyroxine result was obtained on a patient sample on a dimension rxl max with hm instrument.The initial result was reported out to the physician(s), which was questioned.The customer redrew a sample and tested it on the same dimension rxl max with hm instrument, resulting lower.The customer issued a corrected report to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated free thyroxine result.
|