Catalog Number EMAX2PLUS |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/13/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was discovered that the motor device was allowing the irrigation flow to drip.It was reported that the doctor experienced the dripping again using the irrigation system.The reporter did not specify when the issue initially started.It was not reported if there were any delays in the surgical procedure.It was reported that a spare device was available for use.There was patient involvement reported.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device passed all manufacturing specifications.Therefore, the reported condition was not duplicated and confirmed.An assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|