Model Number 10094137 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/09/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
|
|
Event Description
|
It was reported to siemens that a malfunction occurred with the artis zee ceiling system.The customer reported the system was down and unusable as dose was less than 5%.At the time of the event, the system was not being used for a clinical procedure and there is no report of impact to the state of health of any patient involved.
|
|
Manufacturer Narrative
|
Siemens has completed an investigation of the reported event.The investigation confirmed a loss of high voltage stability due to damage or leakage in the welding line of the x-ray window of the tube vacuum housing.In this specific case, the leakage, and the associated arcing progressed slowly so that the user was unable to detecting the issue.The affected x-ray tube has been exchanged as part of the service activity.The manufacturer is not considering further actions resulting from this event.
|
|