Model Number 10094135
Device Problem Electrical Shorting (2926)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/15/2017
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.This event occurred in (b)(6).
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zee floor system.During an acute interventional procedure, a short circuit in the high voltage cable occurred on the cathode side.The patient was safely removed from the system and transferred to an alternative system where the procedure was completed.We are unaware of any impact to the state of health of the patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.During the investigation, it was determined that a hardware issue occurred.The event log showed several warnings of cathode-side arcing from the high voltage generator, resulting in a high voltage error.When high voltage is applied and an error occurs, no radiation is possible.A siemens service engineer replaced the 4 pin high voltage cable and returned the system to clinical operation.No further actions are to be taken as there is no negative awareness in regards to the quality and performance of the affected component.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE FLOOR
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key6445592
MDR Text Key71497994
Report Number2240869-2017-74642
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K123529
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 03/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number10094135
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/15/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N