Model Number 10094135 |
Device Problem
Electrical Shorting (2926)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/15/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.This event occurred in (b)(6).
|
|
Event Description
|
It was reported to siemens that a malfunction occurred while operating the artis zee floor system.During an acute interventional procedure, a short circuit in the high voltage cable occurred on the cathode side.The patient was safely removed from the system and transferred to an alternative system where the procedure was completed.We are unaware of any impact to the state of health of the patient involved.
|
|
Manufacturer Narrative
|
Siemens has completed an investigation of the reported event.During the investigation, it was determined that a hardware issue occurred.The event log showed several warnings of cathode-side arcing from the high voltage generator, resulting in a high voltage error.When high voltage is applied and an error occurs, no radiation is possible.A siemens service engineer replaced the 4 pin high voltage cable and returned the system to clinical operation.No further actions are to be taken as there is no negative awareness in regards to the quality and performance of the affected component.
|
|