The reported event did not occur during a procedure, no patient impact occurred as a result.The amsco examiner 10 lighting system is constructed on a mobile stand and is required to be plugged into a power source to be used.A steris service technician arrived on site, inspected the light, and identified the insulation on the power cord for the light was damaged causing the wire to be exposed and resulting in the reported spark.The technician identified this abrasion on the insulation was most likely caused by the user facility stretching the power cord too far when using the lighting system.The technician replaced the damaged insulation, installed wire ties on the power cord to prevent it from being stretched in the future, tested the lighting system, and confirmed it to be operating according to specification.The technician notified the user facility to refrain from stretching the power cord on the lighting system to prevent future recurrence.No additional issues have been reported.
|