The device referenced in this report was returned for evaluation.S-family hsc confirmed that this issue occurred on vd10c and no evidence could be found that the issue was recurrent or went away after transducer replacement, it is likely that this issue was caused by the initialization issue with the sw vd10x.This issue was investigated via plm defect 483014 and resolved in ve10a, via eco# 649637 released on 02 mar 2017 from plm defect 483014: the root cause of the triple images is an initialization issue when the transducer is selected, which occurs roughly 1 in 1000 probe initializations.The interim solution would be to look for the triple image when the transducer is selected, which can be done by pressing a finger on one end of the transducer, and making sure the image is as expected, i.E.You can see one finger on the image.A review of the device history record (dhr) was performed and there is no information to indicate any non-conformance at the time of manufacturing process.
|