Model Number PM2210
Device Problem Programming Issue (3014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/09/2017
Event Type  malfunction  
Event Description
It was reported that the during a merlin.Net transmission automatic mode switches were noted.It was determined the presence of competitive atrial pacing leading to a device mode switch.While in sensor indicated rate, the atrial events fall into pvarp causing intermittent atrial pacing.Clinician verified that there was no (pmt) -pacemaker mediated tachycardia - no attempt to resolve the competitive atrial pacing, as there have not been anymore alerts.The patient was not symptomatic.The patient would be followed normally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCENT DR RF PACEMAKER
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6460520
MDR Text Key71729830
Report Number2017865-2017-02060
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial
Report Date 04/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2012
Device Model NumberPM2210
Device Lot Number3319186
Other Device ID Number05414734503617
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/09/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/10/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age43 YR
-
-