Model Number C60A3U |
Device Problem
Reset Problem (3019)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/23/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a partial electrical reset.Partial reset occurred on (b)(6) 2017.
|
|
Event Description
|
It was reported that the implantable pulse generator (ipg) device experienced a reset, which resulted in temporary stop of pacing and sensing.An interrogation was performed to clear the reset.The device remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.Performance data collected from the device was received and analyzed.Analysis of the device memory indicated a partial electrical reset.Partial reset occurred on (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was additionally reported that the device was explanted.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|