MAUDE Adverse Event Report: MEDTRONIC, INC. ATTAIN VENOGRAM BALLOON CATHETER; CATHETER, FLOW DIRECTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. ATTAIN VENOGRAM BALLOON CATHETER; CATHETER, FLOW DIRECTED Back to Search Results
Model Number 6215
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/19/2017
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the balloon catheter was returned and analyzed.No anomalies were found.The analyst noted a test guidewire was fully inserted into the balloon catheter.
 
Event Description
It was reported that during the implant procedure, the balloon catheter was inserted into the coronary sinus and as the physician attempted to insert a guidewire into the balloon catheter, it was not possible to enter the lumen of the balloon catheter more than a few centimeters.The balloon catheter was replaced.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN VENOGRAM BALLOON CATHETER
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6470395
MDR Text Key72202741
Report Number2182208-2017-00647
Device Sequence Number1
Product Code DYG
UDI-Device Identifier00613994842114
UDI-Public00613994842114
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K012225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/28/2018
Device Model Number6215
Device Catalogue Number6215
Device Lot Number0061504143
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-