Model Number RT021 |
Device Problems
Crack (1135); Split (2537)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/07/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The rt021 is sold in the usa but has no 510(k) number as it is considered a class i device.We are currently in the process of obtaining further information from the hospital to determine if the complaint rt021 catheter mount caused or contributed to the reported event.We will provide a follow up report upon completion of our investigation.
|
|
Event Description
|
A hospital in (b)(6) reported via a fisher & paykel healthcare field representative that the rt021 catheter mount tubings have "split".This was found before patient use.
|
|
Manufacturer Narrative
|
The rt021 is sold in the usa but has no 510(k) number as it is considered a class i device.Method: six complaint rt021 catheter mounts (two units in unsealed bags and four units in sealed bags) and additional (b)(4) catheter mounts in sealed bags were returned to fisher & paykel healthcare in (b)(4) for investigation, and were visually inspected.In addition, two of the returned complaint units (in sealed bags) were leak tested.Results: visual inspection revealed that the tubing cuff of the returned complaint catheter mounts had split at the connector end.Leak test of the two returned complaint units (in sealed bags) revealed that these units were out of specification.No fault was found with the returned sealed (b)(4) catheter mounts.Conclusion: we are unable to determine the cause of the damage observed on the returned rt021 catheter mounts.All rt021 catheter mounts are pressure tested prior to being released for distribution.Any catheter mount with a split tube would have failed the pressure test.This suggests that the returned catheter mounts were damaged after they were released for distribution.Our user instructions that accompany the rt021 catheter mount state the following: check all connections are tight before use.Perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.
|
|
Event Description
|
A hospital in (b)(6) reported via a fisher & paykel healthcare field representative that the rt021 catheter mount tubings have "split".This was found before patient use.
|
|
Search Alerts/Recalls
|