MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BEQ-HLS 7050 USA
Device Problem Occlusion Within Device (1423)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/20/2017
Event Type  malfunction  
Manufacturer Narrative
The product was disposed by the hospital; therefore a manufacturer laboratory investigation was not possible.A review for similar complaints with a product lab investigation performed was done and the following results could be obtained: the product was visual inspected in the lab of the manufacturer.Some clotted areas were visible on the outlet and inlet side.No clots were rinsed out during cleaning.The product was tested with bovine blood in the laboratory of the manufacturer for its o2, co2 transfer rate as well as for its pressure drop behavior at maximum flow.The product was operating according to the acceptance criteria's and therefore passed the test successfully.Clotting is a known phenomenon to mcp and the cause of this incident was determined to not be attributed to a device related malfunction.Based on these results and the information available at this time, the oxygenator in question operated within mcp specifications.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Event Description
According to the customer: "there was a "shadow" on the cardiohelp oxygenator outlet side.The doctor looked at the shadow and decided to start heparin drip".The sales person told the customer that shadow did appear to be a clot forming and that if the customer had any fear that this would endanger the patient that the customer should change out the hls unit.The product was not exchanged.No patient consequences were reported" (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/BIOLINE COATING
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6471699
MDR Text Key72416188
Report Number8010762-2017-00116
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
PMA/PMN Number
K080592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBEQ-HLS 7050 USA
Device Catalogue Number701052794
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-