Model Number 0693 |
Device Problems
Over-Sensing (1438); Pacing Problem (1439); Ambient Noise Problem (2877); Material Integrity Problem (2978)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/09/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
|
|
Event Description
|
Boston scientific received information that, during the implant of this right ventricular (rv) lead, oversensing of noise led to significant pacing inhibition.Blood was observed in the lead insulation and lead damage was suspected.This rv lead was explanted and replaced prior to pocket closure.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.The complete lead was returned severed in two segments.Visual inspection revealed a cut in the insulation 23 centimeters from the terminal pin.Body fluid/blood was visible in the lumen and was assumed to have entered through the cut.Resistance tests were completed on both segments to assess lead electrical performance.Measurements throughout these tests were within normal limits.Laboratory analysis was able to confirm the field observation of lead damage, but there were no lead characteristics that would have contributed to the reported electrical observations.
|
|
Search Alerts/Recalls
|