Information was received from a consumer via a manufacturer representative regarding a patient who was implanted with a neurostimulator for non-malignant pain.It was reported that the manufacturer representative saw the patient about a month ago and noticed that impedances were elevated around 1,000-2,000 ohms.The patient was seen for normal reprogramming a few times but on the day of the call ((b)(6) 2017) it was the first time that the manufacturer representative noticed the correlation between the high impedances and a lack of stimulation.Impedances were run at 0.7v and noticed the impedances at the bottom of the lead were about 1000 ohms but it started to increase going up the lead.At the top of the lead the impedances were around 3,000 to 4,000 ohms.Having the reference electrode at 0 led to higher impedances than when electrode 7 was the reference electrode.When reference was 0, impedances are low at the top and higher at the bottom but if switched to reference 14 it was the opposite.When the patient was programmed at the bottom of the lead the patient felt stimulation around 3-4v, when the patient was programmed in the middle of the lead they felt stimulation around 5-6v, and when programmed a the tip the amplitude could be turned all the way up and the patient would not feel anything.The patient still had pain and it had never been ¿the way it should be.¿ the patient needed low back coverage but felt stimulation in their legs up to their buttocks.There was a reported fall off of a ladder last year but there was no noticed a change with the stimulator after that fall.The patient had fallen again about a week ago.Electrode impedance testing was done at 3.0v and there were no anomalies found.The following results were recorded from different reference electrodes: reference 0 - around high 2000s to 4000 ohms, reference 7 - all below 2500 ohms, and reference 13 - high 2000s to 3000 ohms.It was noted that these issues had been since implant on (b)(6) 2015.No further complications were reported.
|