MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) CART 9734056 S7 STAFF SHRT 100-120V INTL; SEE H10)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) CART 9734056 S7 STAFF SHRT 100-120V INTL; SEE H10) Back to Search Results
Model Number S7
Device Problems Unintended Arm Motion (1033); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/15/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information was not made available from the site.No procode, common device name and/or 510(k) provided as this device is not released for distribution in the united states.  return requested.Replacement device shipped to site 03/15/2017.No parts have been received by manufacturer for analysis.
 
Event Description
A medtronic representative reported that, while in a cranial biopsy procedure, the site attempted to attach an articulating arm in the aseptic field, however, the support arm was unstable and could not be fixed.The support arm was placed out of aseptic field to be checked and inspected.A second articulating arm was immediately sterilized and brought in to be used instead.No further details regarding the damage, or how it occurred, were provided.Delay in therapy was 5 minutes.The surgeon completed the procedure with the use of the navigation system.There was no impact on patient outcome.
 
Manufacturer Narrative
Additional information: device manufacturing date 12/05/2014.The hardware investigation of the returned articulating arm found that the reported event was related to a mechanical issue.This was a down revision component.The lot code indicates a manufacture date of july 2008.The handle of the arm was locked up with the arm in the released condition.The reported event was confirmed.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CART 9734056 S7 STAFF SHRT 100-120V INTL
Type of Device
SEE H10)
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
amos jarrette
826 coal creek circle
louisville, CO 80027-9710
7208902082
MDR Report Key6481790
MDR Text Key72790933
Report Number1723170-2017-01030
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
SEE H10)
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 05/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9734056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received04/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-