The customer contacted a siemens customer care (ccc) and stated that the integrated multi-sensor technology (imt) sensors were unstable.The customer replaced the imt sensor and ran conditioning.Quality controls were run, which were within acceptable ranges.Siemens regional support center (rsc) and headquarters support center (hsc) specialists reviewed the instrument data and determined that the customer was using the imt sensor for 7 days, whereas siemens claims the use of the imt sensor is for 5 days.Additionally, the customer was spinning the sample tubes for one minute at the speed of 10,000 rotations per minute.The hsc specialist stated that the customer must follow proper tube manufacturer recommendations for centrifugation times and speed.The hsc specialist also recommended that the customer must review proper pre-analytical sample handling procedures including mixing of the sample after phlebotomy to ensure complete dispersion of the anti-coagulant or clot activator throughout the sample.The rsc specialist determined that the customer's dilution factor was lower than expected.The rsc specialist also determined that the customer was not performing the required troubleshooting such as daily maintenance, which must be followed to ensure accuracy and system performance.The cause of the discordant, falsely elevated k result on one patient sample is unknown.The instrument is performing within manufacturing specifications.No further evaluation of device is required.
|
Discordant, falsely elevated potassium (k) results were obtained on one patient sample on a dimension exl 200 instrument.The discordant results were not reported to the physician(s).The sample was repeated on the same instrument, resulting lower and matching the clinical picture of the patient.The repeat result was reported to the physician(s).There are no known reports of patient intervention or adverse health consequence due to the discordant, falsely elevated k results.
|