ETHICON SARL GYNECARE TVT ABREVO CONTINENCE SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC Back to Search Results
Catalog Number TVTOML
Device Problem Difficult to Insert (1316)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Event Description
It was reported that the patient underwent a sling procedure on an unknown date and mesh was used.The white plastic sheath that covers the helical passer would not come off the helical passer.The entire mesh device was removed and a new device was implanted and worked without any difficulty.There were no adverse patient consequences.No additional information will be provided.
 
Manufacturer Narrative
The device received was manipulated and used.The prolene lines, and plastic sheath were cut.Marks of clamps are visible on the plastic needles and plastic sheath.The device was disassembled.Based on the evaluation, this complaint is not a manufacturing issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT ABREVO CONTINENCE SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON SARL
puits-godet 20
neuchatel 2000
SZ  2000
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits-godet 20
neuchatel 2000
SZ   2000
Manufacturer Contact
darlene kyle
route 22 west po box 151
somerville, NJ 08876
9082182792
MDR Report Key6487442
MDR Text Key72657062
Report Number2210968-2017-03339
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2017
Device Catalogue NumberTVTOML
Device Lot Number3916752
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-