Model Number 1258T/86
Device Problems Misconnection (1399); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Device Overstimulation of Tissue (1991)
Event Date 02/15/2017
Event Type  Injury  
Event Description
It was reported that patient received inappropriate therapy due to oversensing on the atrial channel.The physician placed an lv lead into the atrial port.The lead was detecting p-waves well but also detecting far-field events due to the off-label placement.Programming changes were recommended.Patient is in stable condition.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6487910
MDR Text Key72652740
Report Number2938836-2017-21543
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Modification/Adjustment
Type of Report Initial
Report Date 04/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Model Number1258T/86
Device Lot NumberA000019535
Other Device ID Number05414734502610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
(B)(4)
Patient Age71 YR
-
-