Catalog Number 5MAXACE068KIT |
Device Problems
Break (1069); Material Separation (1562); Material Integrity Problem (2978)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/17/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Please note that the following device codes also apply to this complaint: (b)(4).This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
|
Event Description
|
During preparation for a thrombectomy procedure, the hospital staff noticed that the penumbra system ace 68 reperfusion catheter (ace 68) was coming apart upon removal from the packaging of the penumbra system ace 68 hi-flow kit (kit).The damaged ace 68 was noticed prior to use and therefore, was not used in the procedure.The procedure was completed using a new kit.
|
|
Manufacturer Narrative
|
Results: the penumbra system ace 68 reperfuslion catheter (ace 68) was fractured approximately 63.0 cm from the hub and ovalized on the distal tip.Conclusions: evaluation of the returned device revealed that it was fractured.This type of damage typically occurs if the ace 68 is withdrawn from the packaging tray prior to removing the tubing tray.Further evaluation revealed the ace 68 was ovalized on its distal tip.This damage was likely incidental and may have occurred during packaging for return.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Search Alerts/Recalls
|