Model Number 383400 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/10/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Submit date: 13-apr-2017.An investigation is currently underway.Upon completion, the results will be forwarded.
|
|
Event Description
|
The customer states failed motor speed test.No patient involved.
|
|
Manufacturer Narrative
|
Updated with the initial reporter's first name, last and and telephone number.
|
|
Event Description
|
The customer states failed motor speed test.No patient involved.
|
|
Manufacturer Narrative
|
An evaluation of the kangaroo pump was performed for the reported condition of failed motor speed test.The unit was triaged and the customer¿s reported condition was confirmed.A trend has been identified and a corrective and preventative action has been opened to address this issue.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.As necessary.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|