• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Difficult to Insert (1316); Kinked (1339); Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/22/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the patient presented in clinic for follow up when capture thresholds on the lv lead were noted to be inadequate due to lead being dislodged.Lead repositioning was attempted but was not successful due to guidewire not being able to advance down the lead and it was decided to explant the lead.Furthermore, after the lead was explanted a kink near the suture sleeve was observed.The patient did not experience any adverse consequences.
 
Manufacturer Narrative
The damage found was sustained during the surgical procedure.The lead was otherwise normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6491704
MDR Text Key72777152
Report Number2017865-2017-02297
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2019
Device Model Number1458Q/86
Device Lot NumberA000031623
Other Device ID Number05414734503198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age80 YR
-
-