ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) CURRENT DR RF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number 2207-36
Device Problem Difficult to Interrogate (1331)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/23/2017
Event Type  Injury  
Event Description
It was reported that the device was unable to communicate with multiple transmitters.The device had reached the elective replacement indicator and will be replaced.No further information is available at this time.
 
Manufacturer Narrative
An event of no communication was confirmed.During analysis, the device communicated normally via inductive telemetry.Communication could not be established via rf telemetry due to low battery voltage.Longevity estimation was performed and the battery voltage was found to be lower than expected.During the analysis, the device behaved normally and the power consumption of the device was measured and found to be normal.The cause of the premature battery depletion is undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURRENT DR RF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6493864
MDR Text Key72839488
Report Number2938836-2017-21784
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2010
Device Model Number2207-36
Device Lot Number2764896
Other Device ID Number05414734501927
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age73 YR
-
-