BIOMET SPORTS MEDICINE JUGGERKNOT SINGLE LOADED ANCHOR 1.5MM; FASTENER, FIXATION Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Device Or Device Fragments Location Unknown (2590)
Patient Problem No Information (3190)
Event Date 03/16/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It is reported that during a procedure, while the surgeon was trying to insert an anchor, the tip of the inserter fractured.Although, no parts could be found in the patient it is unknown if the fractured piece remained in the patient's bone.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Product was visually examined.The inserter tip has broken off and the shaft is bent at the distal end.The suture/anchor was not returned with the device.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Review of the complaint history determined that no further action is required as no trends were identified.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUGGERKNOT SINGLE LOADED ANCHOR 1.5MM
Type of Device
FASTENER, FIXATION
Manufacturer (Section D)
BIOMET SPORTS MEDICINE
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6494535
MDR Text Key72876674
Report Number0001825034-2017-02576
Device Sequence Number1
Product Code MBI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK110145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number912031
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/27/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-