• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING Back to Search Results
Model Number 4415
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 03/24/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Per field service representative (fsr), he tested delay override and found that the bubble making compressor was not running, therefore ice making compressor was never signaled to turn on.The compressor delay and pressure switch circuit boards were reseated and then the compressor worked properly.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the main (ice making) compressor did not come on after the normal 6 minute delay.The "ice making" light was on but the compressor did not run.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6498688
MDR Text Key73248880
Report Number1828100-2017-00183
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-