BOSTON SCIENTIFIC - MARLBOROUGH RESOLUTION 360¿ CLIP; HEMOSTATIC METAL CLIP FOR THE GI TRACT Back to Search Results
Model Number M00521231
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/28/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.(b)(4).
 
Event Description
It was reported to boston scientific corporation that a resolution 360 clip device was used during a colonoscopy procedure performed on (b)(6) 2017.According to the complainant, during the procedure, subsequent to successful deployment of the clip, the hypotube came off.It was retrieved with a non-bsc forceps, and the procedure was completed at this time.  there were no reported complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Mfg.Site name - (b)(4).Investigation results: visual examination revealed that only the crimped hypotube was returned for analysis.This failure is likely due to anatomical/procedural factors encountered during the procedure which limited the performance of the device.Therefore, the most probable root cause is operational context.A review of the device history record (dhr) was performed and confirmed that this device was manufactured in accordance with the device master record.A labeling review was performed and from the information available, this device was used per the directions for use (dfu) / product label.A search of the complaint database revealed that no similar complaints exist for the specified lot.
 
Event Description
It was reported to boston scientific corporation that a resolution 360 clip device was used during a colonoscopy procedure performed on (b)(6) 2017.According to the complainant, during the procedure, subsequent to successful deployment of the clip, the hypotube came off.It was retrieved with a non-bsc forceps, and the procedure was completed at this time.There were no reported complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTION 360¿ CLIP
Type of Device
HEMOSTATIC METAL CLIP FOR THE GI TRACT
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6500060
MDR Text Key73020821
Report Number3005099803-2017-01087
Device Sequence Number1
Product Code PKL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/13/2018
Device Model NumberM00521231
Device Lot Number0000056088
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-