Model Number M00521231 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/28/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.(b)(4).
|
|
Event Description
|
It was reported to boston scientific corporation that a resolution 360 clip device was used during a colonoscopy procedure performed on (b)(6) 2017.According to the complainant, during the procedure, subsequent to successful deployment of the clip, the hypotube came off.It was retrieved with a non-bsc forceps, and the procedure was completed at this time. there were no reported complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
Mfg.Site name - (b)(4).Investigation results: visual examination revealed that only the crimped hypotube was returned for analysis.This failure is likely due to anatomical/procedural factors encountered during the procedure which limited the performance of the device.Therefore, the most probable root cause is operational context.A review of the device history record (dhr) was performed and confirmed that this device was manufactured in accordance with the device master record.A labeling review was performed and from the information available, this device was used per the directions for use (dfu) / product label.A search of the complaint database revealed that no similar complaints exist for the specified lot.
|
|
Event Description
|
It was reported to boston scientific corporation that a resolution 360 clip device was used during a colonoscopy procedure performed on (b)(6) 2017.According to the complainant, during the procedure, subsequent to successful deployment of the clip, the hypotube came off.It was retrieved with a non-bsc forceps, and the procedure was completed at this time.There were no reported complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|