• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER; NRY Back to Search Results
Catalog Number 3MAXC
Device Problems Break (1069); Material Separation (1562)
Patient Problem No Patient Involvement (2645)
Event Date 03/21/2017
Event Type  malfunction  
Manufacturer Narrative
Results: the penumbra system 3max reperfusion catheter (3max) was fractured approximately 7.0 cm from the hub.Conclusions: evaluation of the returned device revealed that it was fractured.This damage may have occurred due to forceful manipulation of the 3max at extreme angles during insertion into an rhv.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
During preparation for a thrombectomy procedure, the penumbra system 3max reperfusion catheter (3max) completely cracked off about five centimeters from the proximal hub while the physician was attempting to advance it through the rotating hemostasis valve (rhv) on the back table.The damage to the 3max occurred prior to use and therefore, the 3max was not used in the procedure.The procedure was completed using a new catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER
Type of Device
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key6501424
MDR Text Key73122886
Report Number3005168196-2017-00618
Device Sequence Number1
Product Code NRY
UDI-Device Identifier00814548012490
UDI-Public00814548012490
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160449
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date11/30/2017
Device Catalogue Number3MAXC
Device Lot NumberF47012
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-