Model Number 1458Q/86
Device Problem Detachment Of Device Component (1104)
Patient Problem Foreign Body Reaction (1868)
Event Date 03/13/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during the implant procedure, the physician attempted to remove the stylet but the knob broke off and was stuck in the lead.The lead was no longer used and replaced with lead serial number (b)(4).It was noted that after the second lead was placed that the suture sleeve had migrated into the pectoralis tissue and was not visible and was not surgically removed.The patient tolerated the procedure well and would continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6508785
MDR Text Key73336237
Report Number2017865-2017-02455
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model Number1458Q/86
Device Lot NumberA000030881
Other Device ID Number05414734503198
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
(B)(4)
Patient Age63 YR
-
-