A siemens healthcare diagnostics inc.(siemens) customer service engineer (cse) was dispatched to the customer site.The cse aligned and cleaned reagent probes 1 and 2 and primed reagent probes 1 and 2 and sample probe 1 (s1).Service methods were performed.During follow-up visits, the cse replaced the aliquot probe and the integrated multi-sensor technology probe and ran a quick check.The cse cleaned all drains and adjusted vacuum, replaced s1 mixer and probe and performed mixer diagnostics test.Quick check and all functional tests were acceptable.The cse calibrated and ran qc for hemoglobin a1c and performed precision testing on patient samples, all of which were acceptable.The cause of the discordant hemoglobin a1c results is unknown.The instrument is performing according to specifications.No further evaluation of this device is required.
|