Model Number MMT-723NAS
Device Problem Naturally Worn (2988)
Patient Problem Hyperglycemia (1905)
Event Date 03/21/2017
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call of hospitalization for a toe amputation.Customer's blood glucose was 387mg/dl at the time of incident.Customer indicated that the pump was worn in or around an mri machine.Customer was advised to discontinue use of the device and revert to a back-up plan per their doctor's instruction.The customer was also advised that the device would be replaced and agreed to return the product for analysis.No further details given.
 
Manufacturer Narrative
The insulin pump alarmed motor error alarm rewind due to motor encoder signal out of phase.Unable to perform self-test, off no power test, unexpected restart error test, displacement test, prime test, occlusion test and no delivery test due to motor error alarm.However, the insulin pump passed stop current and run current tests.The insulin pump had cracked case at the display window corner and minor scratched display window.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6514476
MDR Text Key73450116
Report Number3004209178-2017-46905
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 05/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-723NAS
Device Catalogue NumberMMT-723NAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received05/06/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization;