The customer contacted the siemens customer care center (ccc).With assistance from the ccc, the customer centered the silver nuts in the reagent probe 1 (r1) and reagent probe 2 (r2), checked the ultrasonics and probes.The customer performed a system check.The check passed.The customer also cleaned the windows and replaced a window that had a cracked frame.After the troubleshooting performed the issue was resolved.The cause of the discordant falsely high hemoglobin a1c result is unknown.The instrument is performing according to specifications.No further evaluation of this device is required.
|