A vcare medium device was given to the surgeon to be tested before use.The balloon would not inflate upon testing, so the device was removed from the sterile field and a new one was introduced.The same error occurred with the second device.A third device was successfully used and the procedure was completed with no harm to the patient.The second device was also removed from the sterile field and both were sent to the clinical engineering department.It was noted that the two failed devices were of the same lot.The device will be returned to the manufacturer for failure analysis and replacement.Hospital will share manufacturer test/evaluation results with medsun.
|