• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number MTO-05400-UCS
Device Problem Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/12/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation at this time.Teleflex will continue to monitor and trend related events.
 
Event Description
The arrow epidural catheter set was used for epidural in labor and delivery.Patient movement in bed resulted in the catheter ripping.The catheter was removed and replaced.The patient's condition was reported as fine.
 
Manufacturer Narrative
(b)(4).A device history record review was performed on the epidural catheter with no relevant findings.The ifu for this kit, (b)(4); rev.3, was reviewed as a part of this complaint investigation.The ifu warns the user, "never tug or quickly pull on catheter during removal from patient to reduce risk of catheter breakage.Do not apply additional tension on the catheter if catheter begins to stretch excessively.Reposition patient to open the vertebral interspaces and re-attempt removal if resistance is encountered or if catheter stretches excessively during removal.During epidural catheter removal, the literature indicates a force of approximately 1/3 of a pound is all that is necessary to exert if patient is properly positioned in the recommended lateral neutral position." a corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed on the epidural catheter with no evidence to indicate a manufacturing related issue.Therefore, the potential cause of the catheter ripping could not be determined based upon the information provided and other remarks: without the sample.
 
Event Description
The arrow epidural catheter set was used for epidural in labor and delivery.Patient movement in bed resulted in the catheter ripping.The catheter was removed and replaced.The patient's condition was reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
312 commerce place
asheboro NC 27203
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key6519382
MDR Text Key73620542
Report Number1036844-2017-00162
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Catalogue NumberMTO-05400-UCS
Device Lot Number23F16M0245
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/17/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-