ARTHREX, INC. BIO-COMPOSITE P-LOCK 2.9 X 15.5 MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
Catalog Number AR-1923BC |
Device Problems
Break (1069); Component Falling (1105)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 03/20/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The devices met all material specifications as received.The evaluation revealed bent and broken tip on one returned device.Device history record review revealed nothing relevant to this event.The complainant's event is typically caused by prying/leveraging the driver.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
|
|
Event Description
|
It was reported that during a shoulder stabilizing procedure, the tip of device broke during surgery and remains in patient's shoulder.Surgeon tried to retrieve fragment, however, was not successful.Surgery was completed using a biofastak, same surgical technique.Copy of surgical report requested, however, not received.No other event details received.
|
|