Catalog Number 72200755
Device Problem Component Falling (1105)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/05/2017
Event Type  malfunction  
Event Description
It was reported that during a a left rotator cuff repair surgery, when first inter anchor was implanted, the press component of shaft fell suddenly,and couldn't be used any more.The anchor had to be removed from the patient.No reported patient injuries.No other complications were noted.
 
Manufacturer Narrative
Visual assessment of the device confirmed the reported complaint of disassembly.The nose cone, sliding ring and spring have come free from the handle, spring was not returned.Examination of the nose cone showed the tabs that interface with the handle have broken off.The condition of the device indicates excessive force was placed on it during use.Potentially the cause for this device failure was excessive forces were applied to the instrument, causing a result in failure.After the evaluation the root cause for the reported issue was determined to be user error.A review of the device history records and complaint files confirmed that no additional complaints have been reported and no abnormalities were noted during the manufacturing process for this lot.There are no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX TI 5.0 ULTRABRAID
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 w. william cannon
austin, TX 78735
MDR Report Key6520589
MDR Text Key73796734
Report Number1219602-2017-00425
Device Sequence Number1
Product Code MBI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K972326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 05/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200755
Device Lot Number50547320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age54 YR
Patient Weight65
-
-