SMITH & NEPHEW, INC. TWINFIX TI 5.0 ULTRABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
Catalog Number 72200755 |
Device Problem
Component Falling (1105)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/05/2017 |
Event Type
malfunction
|
Event Description
|
It was reported that during a a left rotator cuff repair surgery, when first inter anchor was implanted, the press component of shaft fell suddenly,and couldn't be used any more.The anchor had to be removed from the patient.No reported patient injuries.No other complications were noted.
|
|
Manufacturer Narrative
|
Visual assessment of the device confirmed the reported complaint of disassembly.The nose cone, sliding ring and spring have come free from the handle, spring was not returned.Examination of the nose cone showed the tabs that interface with the handle have broken off.The condition of the device indicates excessive force was placed on it during use.Potentially the cause for this device failure was excessive forces were applied to the instrument, causing a result in failure.After the evaluation the root cause for the reported issue was determined to be user error.A review of the device history records and complaint files confirmed that no additional complaints have been reported and no abnormalities were noted during the manufacturing process for this lot.There are no indications that would suggest that the device did not meet product specifications upon release into distribution.
|
|