ADVANCED STERILIZATION PRODUCTS STERRAD® CYCLESURE® 24 BIOLOGICAL INDICATOR; INDICATOR, BIOLOGICAL (FRC)
|
Back to Search Results |
|
Catalog Number 14324 |
Device Problems
Chemical Problem (2893); Device Operates Differently Than Expected (2913)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/04/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The correct brand name is sterrad® sealsure chemical indicator tape.The correct common device is indicator, chemical.The correct catalog number is 14202_90.
|
|
Event Description
|
A customer reported the sterrad® sealsure chemical indicator tape did not change color correctly after a completed sterrad® 100s cycle.It is unknown if the affected load was reprocessed or released for use on patients, however, there have been no reports of infection, injury or harm to patient(s) associated with this issue.Although there is no report of patient injury or harm and no prior incidents have resulted in serious injury, advanced sterilization products (asp) has determined in this situation sterility cannot be assured.Therefore, as a matter of policy asp had decided to report all incidents of sterrad® sealsure chemical indicator tape not changing color correctly.
|
|
Manufacturer Narrative
|
(b)(4).Correction: information provided by the customer indicates no load was affected because no load was involved in this cycle.Asp investigation summary: the investigation included a review of the device/batch history record review (dhr), trending analysis by lot number, visual analysis, retains analysis, concomitant product evaluation, and system risk analysis (sra).The dhr/batch record review concluded all process specifications were met before the release of the product.Trending analysis by lot number was reviewed from 10/07/2016 to 04/05/2017 and trending was not exceeded.Visual analysis was not performed as the product was not returned for evaluation since user error was determined to be the cause of the issue.Retains testing was not performed since user error was determined to be the cause of the reported issue.An evaluation of the concomitant sterrad® 100s was not performed since user error was determined to be the cause of the reported issue the sra indicates the risk associated with a quality problem with no impact on safety is "low." the assignable cause of the issue can be attributed to incorrect placement of the sealsure tape.The customer reported a portion of the chemical indicator (ci) tape was placed facing down on the sterrad® 100s unit rack.Technical support representative (tsr) educated the customer on proper chemical indicator tape placement and advised the customer to re-run another cycle and ensure the tape was fully exposed, not covered by overlapped tape, wrap or other load items.The issue will continue to be tracked and trended.
|
|
Search Alerts/Recalls
|
|
|