• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHARTER MEDICAL CHARTER MEDICAL CRYOBAG; CRYOBAG STORAGE CONTAINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHARTER MEDICAL CHARTER MEDICAL CRYOBAG; CRYOBAG STORAGE CONTAINER Back to Search Results
Lot Number 142707
Device Problems Fluid/Blood Leak (1250); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2017
Event Type  malfunction  
Event Description
It was reported that during thaw of a patient product, the bag integrity was compromised, accounting for an approximate 10ml loss of an 80ml product.The bag was removed to the bmt cell therapy lab for recovery.Investigation of the break showed the seam between the port and the bag broke.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHARTER MEDICAL CRYOBAG
Type of Device
CRYOBAG STORAGE CONTAINER
Manufacturer (Section D)
CHARTER MEDICAL
MDR Report Key6521803
MDR Text Key73825778
Report NumberMW5069315
Device Sequence Number1
Product Code LPZ
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Attorney
Type of Report Initial
Report Date 04/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot Number142707
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-