• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ESSENCE; COMPRESSOR, AIR, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ESSENCE; COMPRESSOR, AIR, PORTABLE Back to Search Results
Model Number 1095060
Device Problem Cut In Material (2454)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/22/2017
Event Type  malfunction  
Event Description
It was reported that this device had no air flow.On receipt of the device for evaluation it was observed that the main cord was severed leading to exposed wires.It appears as though the cord was intentionally cut, and the customer was approached with regard to confirming this assertion.They have not provided any information regarding how the device came to be in this condition, only that it was in this condition when they received it from their customer.The ifu for this device contains the warning 'the power supply cord cannot be replaced by the user.In case the power supply cord becomes damaged, contact philips respironics customer service for replacement (there is no service option)'.No patient harm or death has been reported and the device is not life-sustaining/supporting.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ESSENCE
Type of Device
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, west sussex PO20 2FT
UK  PO20 2FT
Manufacturer Contact
alessandro agosti
chichester business park
city fields way, tangmere
chichester, west sussex PO20 -2FT
UK   PO20 2FT
704231549
MDR Report Key6525677
MDR Text Key74046186
Report Number9681154-2017-00006
Device Sequence Number1
Product Code CAF
UDI-Device Identifier00383730001821
UDI-Public00383730001821
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042655
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 03/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1095060
Device Catalogue Number1100312
Device Lot Number131011
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-