The customer contacted the siemens customer care center (ccc).A ccc specialist reviewed the instrument data which indicated that quality control (qc) was out of specification high for level 1 but qc level 2 was not affected.A siemens customer service engineer (cse) was at the customer site for a separate issue.The cse power flushed the integrated multi-sensor technology (imt) system and aligned it.The cse performed dilution check and ran electrolytes qc, which passed.The customer ran patient comparison, which resulted satisfactory.The cause of the discordant, falsely low na results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
Discordant, falsely low sodium (na) results were obtained on two patient samples on a dimension vista 1500 instrument.The discordant results were not reported to the physician(s).The samples were repeated on an alternate dimension vista instrument, resulting higher.The repeated results were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low na results.
|