• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ATRICURE, INC. COBRA FUSION 150 ABLATION MINIMALLY INVASIVE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRICURE, INC. COBRA FUSION 150 ABLATION MINIMALLY INVASIVE SYSTEM Back to Search Results
Model Number 001-700-001MI
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/30/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4) the device has not yet been received by atricure.When additional information is received a supplemental report will be submitted.
 
Event Description
During a off-pump cardiac ablation procedure, the magnet from the fusion magnet retriever got loose at the end of the procedure.The case was completed without any delay and patient outcome was not affected.
 
Manufacturer Narrative
(b)(4).The device was returned to atricure and pursuant to (b)(4) and was visually and functionally tested.The magnet housing on the swivel positioner had failed allowing the magnet to become dislodged from the housing.The magnet was not returned with the device.The complaint was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBRA FUSION 150 ABLATION MINIMALLY INVASIVE SYSTEM
Type of Device
COBRA FUSION 150 ABLATION MINIMALLY INVASIVE SYSTEM
Manufacturer (Section D)
ATRICURE, INC.
7555 innovation way
cincinnati OH 45040 0000
Manufacturer Contact
ranjana iyer
7555 innovation way
cincinnati, OH 45040-0000
5137555328
MDR Report Key6527730
MDR Text Key73908292
Report Number3011706110-2017-00042
Device Sequence Number1
Product Code OCL
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
K113475
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/01/2019
Device Model Number001-700-001MI
Device Catalogue Number001-700-001MI
Device Lot Number66132
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
-
-