Catalog Number CAT02462
Device Problem Cut In Material (2454)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device manufacture date is not known at this time because lot number could not be confirmed however, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported the suture tape cut through the anchor.
 
Manufacturer Narrative
The product was not returned for investigation therefore the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: suture tape cut through the anchor probable root cause: "design: anchor design does not facilitate suture adjustment.Anchor barbs too aggressive for suture.Anchor does not have adequate bone engagement feature.Manufacturing: anchor not manufactured to specification.Application: excessive force applied to suture.Wrong suture type used.Removal of inserter before suture adjustment.Anchor seated too deep, suture cut".The device manufacture date is not known.(b)(4).
 
Event Description
It was reported the suture tape cut through the anchor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CINCHLOCK SS KNOTLESS ANCHOR
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
victoria milich
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key6534841
MDR Text Key74255131
Report Number0002936485-2017-00413
Device Sequence Number1
Product Code MBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCAT02462
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/06/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-