• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CINCINNATI SUB-ZERO, LLC. HEMOTHERM 400 CE; CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CINCINNATI SUB-ZERO, LLC. HEMOTHERM 400 CE; CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER. Back to Search Results
Model Number 400CE
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/27/2017
Event Type  malfunction  
Manufacturer Narrative
Csz medical technical support received a complaint stating the hematherm pump was not working.The user facility had a backup device that allowed the procedure to continue successfully.Csz 3rd party technician (b)(6) found a loose connection between the f6 fuse and the fuse holder.The technician secured the fuse to resolve the issue.
 
Event Description
Cincinnati sub-zero medical technical support received a complaint stating the hematherm pump was not working.The user facility had a backup device and the procedure was able to continue.There was no patient or user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOTHERM 400 CE
Type of Device
CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER.
Manufacturer (Section D)
CINCINNATI SUB-ZERO, LLC.
12011 mosteller road
cincinnati OH 45241
Manufacturer Contact
christina miracle
12011 mosteller road
cincinnati, OH 45241
MDR Report Key6538239
MDR Text Key74406861
Report Number1516825-2017-00012
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122813
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number400CE
Device Catalogue Number86022
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-