• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. INTERNAL BRACE LIGAMENT; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. INTERNAL BRACE LIGAMENT; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number AR-1688-CP
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/11/2017
Event Type  Injury  
Manufacturer Narrative
Patient demographics (date of birth, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested for evaluation but was not returned therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.Complainant's event is typically caused by not inserting the implant co-axial to the bone tunnel or improper bone preparation.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
 
Event Description
It was reported that during an fdl transfer in conjunction with the internal brace for spring ligament, the surgeon was using an internal brace ligament kit.Upon implanting the 3.5 mm swivelock from the internal brace kit, the driver disconnected before the anchor was fully seated; leaving 1-2 threads proud.The proud portion of the anchor was removed with a rongeur and the case was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERNAL BRACE LIGAMENT
Type of Device
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath, sr mdr analyst
1370 creekside boulevard
naples, FL 34108-1945
8009337001
MDR Report Key6539931
MDR Text Key74278743
Report Number1220246-2017-00155
Device Sequence Number1
Product Code MAI
UDI-Device Identifier00888867096080
UDI-Public00888867096080
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101823
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/01/2018
Device Catalogue NumberAR-1688-CP
Device Lot Number10082955
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age34 YR
-
-