• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG PLAINFIELD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG PLAINFIELD Back to Search Results
Model Number 383401
Device Problem Power Problem (3010)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.The unit was sent to a service center for repair.Upon triage, a technician found that on the returned power cable there was a point on the cable where the internal wiring was exposed.The power source was replaced and the unit passed recertification without problems.The unit was restored to proper operation.
 
Event Description
It was reported to covidien on (b)(6) 2017 that an issue occurred with an enteral feeding pump.Upon triage on 06-apr-2017 the technician has found an exposed copper wire on the power cord.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Manufacturer (Section D)
COVIDIEN MFG PLAINFIELD
2824 airwest blvd
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN MFG PLAINFIELD
2824 airwest blvd
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key6543004
MDR Text Key74498559
Report Number3008361498-2017-05112
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 04/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date03/16/2014
Device Model Number383401
Device Catalogue Number383401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2017
Date Manufacturer Received04/06/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-