LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS, Back to Search Results
Model Number 16-02-80
Device Problem Biofilm coating in Device (1062)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/04/2017
Event Type  malfunction  
Manufacturer Narrative
There is no known patient involvement.Pma510(k): the heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Recall number: livanova (b)(4) implemented a field safety notice for disinfection and cleaning of sorin heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).Through follow-up communication with the customer, livanova (b)(4) has been informed that no mycobacterium chimaera was found in the water samples.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.Device not yet returned.
 
Event Description
Livanova (b)(4) received a report that biofilm was identified in the tank of the heater-cooler system 3t when the unit was opened up to perform an internal hose replacement during maintenance.This issue was discovered by a livanova field service representative during routine service.There is no known patient involvement.
 
Manufacturer Narrative
Through follow-up communication with the customer, (b)(4) learned that the unit is not in use and that the facility will plan a deep disinfection.The nature of the biofilm is not known and no other sample are available for further investigation.The root cause in this case could not be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS,
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6546338
MDR Text Key74492614
Report Number9611109-2017-00340
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Remedial Action Notification
Type of Report Initial,Followup
Report Date 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number16-02-80
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/24/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/29/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015.
Patient Sequence Number1
-
-