• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SA ROSA SURGICAL DEVICE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SA ROSA SURGICAL DEVICE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number N/A
Device Problem Calibration Problem (2890)
Patient Problem No Patient Involvement (2645)
Event Date 04/13/2017
Event Type  malfunction  
Manufacturer Narrative
The optical distance sensor of ro10011 device has not been evaluated yet for investigation purpose.Once the evaluation will be performed, a follow-up medwatch report will be submitted.
 
Event Description
During device routine maintenance performed by a medtech field service engineer, the optical distance sensor failed the accuracy checks.
 
Manufacturer Narrative
The distance sensor was inspected and it was pointed out that it was manufactured according to a previous version of drawing, that use a circular hole for the pin.The design has since evolved and now uses an oblong hole for the pin to avoid similar incident and facilitate the mounting of instruments on the interface block.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA SURGICAL DEVICE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key6547529
MDR Text Key74592345
Report Number3009185973-2017-00606
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Replace
Type of Report Initial,Followup
Report Date 01/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received01/30/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-