The customer contacted a siemens customer care center (ccc) specialist.The ccc reviewed the data supplied.The ccc found the customer's quality control (qc) was out of range at the time of the event.The customer replaced the v-lyte diluent.The customer ran qc, resulting in range.The customer repeated the same samples, resulting within the expected clinical range.A siemens headquarters support center (hsc) reviewed the event.Hsc confirmed the reported issue.The cause of the discordant, falsely elevated sodium result is due to insufficient v-lyte diluent to perform the integrated multisensor technology dilution for patient samples as a result of the v-lyte diluent bottle not being properly seated by the user.The instrument is performing according to specifications.No further evaluation of the device is required.
|
A discordant, falsely elevated sodium result was obtained on a patient sample on a dimension vista 500 instrument.The initial result was released to the physician(s).The customer repeated the same sample on an alternate dimension vista instrument, resulting lower.The customer issued a corrected report with the alternate instrument result.There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated sodium result.
|