• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 050-87215
Device Problems Fluid/Blood Leak (1250); Torn Material (3024)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/15/2017
Event Type  malfunction  
Manufacturer Narrative
Unique identifier (udi): (b)(4) - the device was not returned to the manufacturer for physical evaluation, and the lot number was not able to be obtained to date.Distribution records were reviewed to identify potential lots of this product shipped to the customer over the past 3 months.The entire set of lots have been sold and distributed.Batch records for the lots identified were reviewed and confirmed there were no deviations or non-conformances during the manufacturing process.The device history record [dhr] of this product was reviewed and no nonconformance reports or other abnormalities during the assembly of this lot were found.In addition, the batch record review confirmed the labeling, material, and process controls were within specification.
 
Event Description
It was reported by patient's peritoneal dialysis registered nurse (pdrn) that the cycler alarmed for air detected in cassette warning last night during treatment.Patient's pdrn reported that upon opening the cassette door on the following morning, the cassette lining had torn and fluid began to leak.The patient's pdrn later confirmed that the fluid leak did not result in any adverse events.Pdrn performed manual drain and cancelled the treatment.Additional information was requested but was not available.The set was discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE PATIENT CONN
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX   88780
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key6555673
MDR Text Key74791718
Report Number8030665-2017-00231
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Catalogue Number050-87215
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received04/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DELFLEX PD FLUID; LIBERTY CYCLER
-
-