ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
|
Back to Search Results |
|
Catalog Number 050-87215 |
Device Problems
Fluid/Blood Leak (1250); Torn Material (3024)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/15/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Unique identifier (udi): (b)(4) - the device was not returned to the manufacturer for physical evaluation, and the lot number was not able to be obtained to date.Distribution records were reviewed to identify potential lots of this product shipped to the customer over the past 3 months.The entire set of lots have been sold and distributed.Batch records for the lots identified were reviewed and confirmed there were no deviations or non-conformances during the manufacturing process.The device history record [dhr] of this product was reviewed and no nonconformance reports or other abnormalities during the assembly of this lot were found.In addition, the batch record review confirmed the labeling, material, and process controls were within specification.
|
|
Event Description
|
It was reported by patient's peritoneal dialysis registered nurse (pdrn) that the cycler alarmed for air detected in cassette warning last night during treatment.Patient's pdrn reported that upon opening the cassette door on the following morning, the cassette lining had torn and fluid began to leak.The patient's pdrn later confirmed that the fluid leak did not result in any adverse events.Pdrn performed manual drain and cancelled the treatment.Additional information was requested but was not available.The set was discarded.
|
|