Model Number CD3357-40Q
Device Problem False Alarm (1013)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/14/2017
Event Type  malfunction  
Event Description
It was reported that during a generator change out after defibrillation threshold successful testing, an alert appeared stating, "possible high voltage circuit damage" with an out of range hv lead impedance.Capacitor maintenance charge was successfully performed, and the alert did not re-appear.A false alert most likely due to the cautery was noted.Firmware download was successfully performed, and the alert was cleared.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6555879
MDR Text Key74790267
Report Number2938836-2017-23852
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Model NumberCD3357-40Q
Device Lot NumberA000035264
Other Device ID Number05414734508216
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/03/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
-
-