The customer did not provide the patient sample for in-house testing.A siemens headquarters support center (hsc) specialist reviewed the information provided by the customer.As per immulite 2000 xpi instructions for use, "as with any immunoassay test on the immulite 2000 system results obtained from the assay should always be used in combination with the clinical examination, patient medical history, and other findings".The hsc specialist stated that the differences in results could be due to different assay standardizations, the use of different antibodies between the two assays, or the difference in assay architectures.The cause of the discordant, falsely low anti-tg ab results on one patient sample is unknown.The device is performing within manufacturing specifications.No further evaluation of device is required.
|
Discordant, falsely low autoantibodies to thyroglobulin (anti-tg ab) results were obtained on one patient sample upon initial and repeat testing on an immulite 2000 xpi instrument.The reagent lot was not provided by the customer.The sample was sent to an alternate laboratory where it was tested on an alternate platform, resulting higher.The initial result was reported to the physician(s) and questioned.It is unknown if the result obtained on the alternate platform was reported to the physician(s).There are no reports of patient intervention or adverse health consequence due to the discordant, falsely low anti-tg ab results.
|